APT3700N Smart Pressure Transmitter for Nuclear Service Operation Manual M3700N-K01E
49
±0.005% of Span per Volt
Static Pressure Effects
ZERO : ±0.1% of URL Per 7MPa
SPAN : ±0.2% of reading per 7MPa
Mounting Position Effects
ZERO Shift up to 350Pa No SPAN Effect
6.6. Functional Specification
Service Liquid, gas
Output
Analog 4 to 20 mA dc and Digital HART Communication
Power supply
12.0 ~ 45 Vdc -- operation
17.4 ~ 45 Vdc -- HART Communications
SPAN and ZERO
Using a button (Non-Safety Related Equipment)
HART Communication (Safety Related Equipment)
Temperature Limit
Operation Temp. : -40 ~ +85
℃
(Non-LCD)
-30
~
+80
℃
(LCD)
Process Temp. : -40 ~ +120
℃
Humidity Limit
5 ~ 100% RH (IP67)
Turn-On Time
MAX. 3 seconds
Response Time
: 200 ms
Pressure Range