APT3700N Smart Pressure Transmitter for Nuclear Service Operation Manual M3700N-K01E
45
5.9. LCD Code
Message
Description
Remarks
ADJ-U
While performing with the Zero Adj button, Zero leaves the
set territory(Upper Part)
ADJ-L
While performing with the Zero Adj button, Zero leaves the
set territory(Lower Part)
ZERO
Zero Button is pressed
SPAN
Span Button is pressed
BT-ERR
Button Pressing Sequence Error
P-LOCK
Protect is Locked in the state of Button Use Error
ZT-ERR
Zero Trim range of Error is Produced (10%)
-TR-
Zero Trim Finished
ZR-ERR
Zero button system Range of Performance Error
SP-ERR
Span button system Range of Performance Error
-ZR-
Zero button system Finished
-SP-
Span button system Finished
-ZA-
Zero Adjustment Finished
-DONE-
Finished Using button of changing other settings
RNGOVR
Using button of changing other settings when going over
input scale
LCD_OV
Shown number of LCD expression possibility going over
the input scale
SCD-ER Sensor
Code
Error
F-RST
Flash system data Reset
F-LOCK
While Flash system data Reset, the Protect is Locked.
F-FAIL
Flash system data Reset is failed
-FR- Flash
Reset
Finished