English–137M7122 DPI 705E / DPI 705E IS User Manual | 31
4. Press the
key 7
times to navigate to the “
” (Clock) screen as shown below.
5. Press the
key to enter the clock setting option.
6. The current set time is displayed in a “
hh.mm
” format as shown below. 24-hour format is
used.
7. To change the time value, press the
key to increment the current (flashing) digit, and
the
key to move to the next digit.
8. Press the
key to save the changes and return to the previous screen (or press the
key to return without saving).
9. Press the
key to return to the sensor reading screen.
3.2.8 Custom Unit
This setting allows a scaling factor (conversion factor) to be set for the pressure measurement
and this will be saved as a Custom Unit. The scaling factor set will be applied automatically for
all existing units in the instrument. The scaling factor will only be applied when the custom unit is
selected during pressure measurement.
To set a custom scaling factor: