Copyright 2019 Baker Hughes Company.
28 | 137M7122 DPI 705E / DPI 705E IS User Manual–English
1. Ensure the display is on the sensor reading screen.
2. Press the
,
and
keys simultaneously.
3. The display will show “
” (Advanced) which indicates that the unit is now in the
Settings
option.
4. Press the
key 5
times to navigate to the “
” (Battery) screen as shown below.
5. Press the
key to enter the battery type setting option.
6. The default battery type is Alkaline displayed as “
” (Battery.Alkaline) shown below.
7. To change the battery type press the
key. This will show the next battery type
indicated by “
” (Battery.Nickel) for Nickel Metal Hydride or “
” (Battery.Lithium) for
Lithium.
8. Press the
key to save the desired battery type and return to the previous screen.
9. Press the
key to leave the screen without saving the changes.
10. Press the
key to return to the sensor reading screen.
3.2.6 Date
This setting changes the system date.
To change the date: