3. The display will show “
” (Advanced) which indicates that
Settings
function is selected.
4. Two types of operations are available:
a. Go to Advanced Settings
There are 7 advanced settings which can be accessed by pressing the
key on
the “
” (Advanced) screen. A PIN will be requested before to gain access to these
settings.
b. Go to General Settings
There are 9 general settings which can be accessed by pressing the
key from the
“
” (Advanced) screen.
The sections below explain the two categories of settings
3.1 Advanced Settings
These features can be accessed through the Advanced Settings which requires a 4-digit unlock
PIN.
3.1.1 Adjust Sensor Calibration
This feature adjusts the calibration of the internal or external sensor. The same procedure is used
for the internal and external sensor. The display shows “
” (Internal Sensor) when
calibrating the internal sensor and “
” (External Sensor) for the external sensor.
Note:
All sensors allow a two point calibration only.
It is recommended to set calibration point “
” (Calibration Point
1) near zero and “
”
(Calibration Point
2) near the full-scale value of the sensor.
Setting
Description
Adjust Sensor Calibration Perform calibration of the sensor
Set Calibration Date
Change calibration date
Set Calibration Due
Interval
Change calibration interval (default 365 days)
Change User PIN
Change user PIN or lock code
Units Lock
Disable or enable measurement units (all units are
enabled by default)
Lock Display
Lock user interface to allow restricted access to
certain functionality
Factory Reset
Reset the instrument to its factory settings
INFORMATION
The default PIN is 4321. It is highly recommended to change
the default PIN for security reasons. Instructions to change the user PIN are
defined in Section