3.1.6 Units Lock
This feature allows certain units that are not required to be disabled. By default, all units are
enabled. To disable units or re-enable previously disabled units:
1. Ensure the display is on the sensor reading screen.
2. Press the
,
and
keys simultaneously.
3. The display will show “
” (Advanced) which indicates that the settings option is selected.
4. Press the
key to enter
Advanced Settings
.
5. Enter PIN. The steps to enter PIN are indicated in above Section
6. If the PIN is correct, the display will show “
” (Adjust).
7. Press the
key 4
times to navigate to the “
” (Units) screen as shown below.
8. Press the
key to enter the Units setting screen.
Note:
The screen shows the current measurement unit and indicates its state with text “ ”
(On) - enabled and “
” (Off) - disabled.
9. To disable the unit, press the
key.
10. Press the
key to go to the Next Unit.