4. Press the
key 4
times to navigate to the auto shutdown screen. “
” (On.OFF) will
be displayed as shown below.
5. Press the
key to enter the auto shutdown setting option.
6. The display will then show the default or the last saved auto shutdown value. Default option
is “
” (Off.10) which means 10
minutes timeout on no activity. The second option is
“
” (Off.no) which means the auto shutdown will be disabled.
7. To toggle the options, press the
key.
8. To accept the value and return to the previous screen, press the
key.
9. To leave the screen without saving, press the
key.
10. Press the
key to return to the sensor reading screen.
3.2.5 Battery Type
The battery type setting allows configuration of the fitted battery type to get maximum battery life
and get accurate capacity readings.
The DPI
705E supports four battery types:
•
Alkaline (Al)
•
Lithium (LiFeS)
•
Nickel Cadmium (Ni
-
Cd)
•
Nickel Metal Hydride (Ni
-
MH).
The latter two are both Nickel based chemistry and have a single setting type “
”
(Battery.Nickel).
Note:
The Intrinsically Safe DPI
705E
-
IS can only be used with the first two battery technologies
listed (Al and LiFeS) - refer to safety manual.
To change the battery type: