English–137M7122 DPI 705E / DPI 705E IS User Manual | 21
11. Finally, once all units are configured, press the
key to save the settings.
12. Press the
key to return to the
Advanced Setting
screen.
13. Press the
key again to return to the sensor reading screen.
3.1.7 Lock Display
This feature allows restricted access to certain features of the instrument.
To lock or unlock the instrument display:
1. Ensure the display is on the sensor reading screen.
2. Press the
,
and
keys simultaneously.
3. The display will show “
” (Advanced) which indicates that the settings option is selected.
4. Press the
key to enter
Advanced Settings.
5. Enter PIN. The steps to enter PIN are indicated in above Section
6. If the PIN is correct, the display will show “
” (Adjust).
7. Press the
key 5
times to navigate to the “
” (Lock) screen as shown below.
8. Press the
key to enter the Lock setting screen.
9. The default Lock /Unlock setting will be displayed as “
” (Off) which means the lock is
disabled.
10. To enable the lock, press the
key. The display will then show “ ” (On) which means
lock is enabled.
INFORMATION
Enabling the display lock will prevent certain functions from
being performed such as changing units, setting the alarm, filter, tare etc.