18
DL-00096 Rev. 8
Alarm Message
Replace C-filtr
Replace P-filtr
Wash I-filtr
Replace batt
XX: Service*
*Value: 71-99
Description
Compressor filter needs
replacement.
Patient filter needs
replacement.
Input filter needs to be
cleaned.
Main battery has exceeded
300 recharge cycles.
Service required.
Action
Replace filter. Press and
hold mute key for more
than 4 seconds to reset
the hour counter.
Replace filter. Press and
hold mute key for more
than 4 seconds to reset
the hour counter.
Wash input filter. Press
and hold mute key for
more than 4 seconds
to
reset the hour counter.
Contact your distributor
for replacement.
Note: This alarm will be
automatically cleared once
a new battery is inserted.
Battery will continue to
function after 300 recharge
cycles, but for a shorter time
than a new battery.
Contact your distributor.
Low Priority Alarms
Other Messages
Message
Power removed
SW Ver N.NNN
Self test
Shutting off
External power
Description
External power has been
disconnected; unit is now
running on battery.
Display of software version
during startup and calibra
-
tion. This includes test of
beeper and LEDs.
Indicates unit is going
through its test at startup.
Displayed for approximately
10 seconds.
Displayed while unit goes
through its power-down
sequence.
External power is
connected.
Action
No action is required.
No action is required.
No action is required.
No action is required.
No action is required.