17
DL-00096 Rev. 8
Alarm Message
Check Vents
Low Battery
XX: Service!*
*Value: 21-50
Description
Device is unable to maintain
oxygen purity.
Estimated battery life less
than 17 minutes.
Service required.
Action
Be sure air inlet/outlet
has not been blocked. If
alarm persists, contact
your distributor.
Charge the battery pack
by plugging in to power
supply.
Note: The message will be
automatically cleared when
plugged in to power supply.
Contact your distributor.
High Priority Alarms
Note: These alarms will allow the Portable Oxygen Concentrator to continue operating.
Alarm Message
Check Cannula
XX: Service*
*Value: 51-70
Description
No breath detected for 5
minutes.
Service required.
Action
Check the cannula con-
nection. Be sure to
breathe through nose. If
alarm persists, contact
your distributor.
Note: The message will be
automatically cleared when
breathing is detected.
Contact your distributor.
Medium Priority Alarms