23
DL-00096 Rev. 8
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electric fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
IEC 60601 test
level
± 6kV contact
± 8kV air
± 2kV for
power supply
lines
± 1kV for
input/output
lines
± 1kV
differential mode
± 2kV common
mode
Electromagnetic
environment/guidance
Floors should be wood,
concrete, or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should
be at least 30%.
Mains power quality
should be that of a typical
commercial or hospital
environment.
Mains power quality
should be that of a typical
commercial or hospital
environment.
Guidance and manufacturer’s declaration: electromagnetic immunity
The Portable Oxygen Concentrator is intended for use in the electromagnetic
environment specified below. The customer or the user of the concentrator should
assure that it is used in such an environment.
Compliance
level
± 15kV contact
± 8kV air
± 2kV for
power supply
lines
± 1kV for
input/output
lines
± 1kV
differential mode
± 2kV common
mode