7. Technical Specifications
I t e m
Sp e c i f i c a t i o n s
Power Supply
100-120VAC, 50/60Hz, 0.5A (For 120V system)
220V-240VAC, 50/60Hz, 0.3 A (For 230V system)
Therapy Pressure
4 –18 cmH
2
O (adjustable in 0.5 cmH
2
O increment)
Ramp Time
0 – 45 minutes (adjustable in 5-minute increment)
Ramp Starting Pressure
3 – 17 cmH
2
O (adjustable in 0.5 cmH
2
O increment)
Dimensions (W x D x H)
16.5 X 24.5 X13.7 cm
Weight
1.4 kg (3.1 lbs)
Sound Level
30 dBA at 10 cmH
2
O, 1 meter distance
Water Capacity:
350ml
Heater Settings:
1 to 6 from 30
°
C-60
°
C
Pressure Drop:
0.2cmH
2
O @ 60LPM
Humidity Output:
10-40mgH
2
O/L
Temperature Operating:
+5
°
C to +35(C (+41°F to +95°F)
Storage: -15
to 50
(+5°F to +122°F)
Shipping: -15
to 70
(+5°F to +158°F)
Environment
Humidity
Operating: 15%RH to 95%RH non-condensing
Storage: 10%RH to 90%RH non-condensing
Shipping: 10%RH to 90%RH non-condensing
Air Tubing
Flexible plastic, 1.8m (approx.)
EN 60601-1
EN 60601-1-2
Classification:
Class I, Type BF, IPX0
Applied Parts Nasal Mask
Not suitable for use in the presence of a flammable anesthetic
mixture (No AP/APG Protection)
Continuous operation.
Note:
the manufacturer reserves the right to modify the specification without notice.