5.2 Tubing and Mask
The tubing and mask should be checked and cleaned daily. Please refer to the cleaning
instruction packaged with the accessories.
1. Disconnect the air tubing from the air outlet of the device.
2. Remove the air tubing and headgear straps from the nasal mask.
3. Use a mild detergent and prepare the detergent according to manufacturer's recommendations.
Tubing and mask can be washed and rinsed in warm water.
4. Rinse thoroughly and make sure they are completely dry before next use.
5. All items of the mask and air tubing are subject to normal wear and tear and may eventually be
replaced. Replace the mask and the air tubing if they are damaged.
WARNING:
Do not use any cleaner containing fragrance or conditioners as they will
leave a residue.
WARNING:
The mask must not be re-used by another person. This is to avoid the risk of
cross-infection.
5.3
Air Filter
The air filter should be checked and replaced every 30 days, or more often if this device is
operated in a dusty environment.
1.
Remove the dirty filter from the enclosure on the front
of the device.
2.
Insert a new filter.
5.4
Water Chamber
1.
Turn the device off and allow the heater and water to cool.
2.
Disconnect the flexible tubing from the water chamber. Slide the water chamber out of the
heater platform. Empty the remaining water.
3.
Push the water chamber slider to the left and gently remove
the chamber base by pressing downward on the metal base
tab.
4.
Use a mild detergent to wash all the chamber parts. Rinse all
the parts with clear water and allow them to air dry.
5.
All items of the chamber are subject to normal wear and tear
14