Additional instruction for Physician and Technician
(Do not distribute to patients)
1. To Set the Therapy Pressure
1.
Press
"MENU"
to select
[P XX.XcmH
2
O]
menu while in the standby screen.
2.
Hold the
"UP"
and
"DOWN"
button, and then simultaneously press the
"MENU"
button for
one second. Meanwhile, the LCD screen
[P XX.XcmH
2
O]
should start blinking to allow
you to adjust the therapy pressure from 4 to 18 cm H
2
O.
3.
Press
"UP"
or
"DOWN"
button to increase or decrease the pressure setting. The increment
or decrement of pressure setting is 0.5 cm H
2
O.
4.
After selecting the prescribed pressure, press
"MENU"
to confirm.
5.
Press
“START/STANDBY”
button to go back to standby screen or leave the device to
automatically go back to standby screen 5 seconds later.
2. To Re-set the Compliance Meter & Total Meter
1.
Press
"MENU"
to select
[TM XXXX.X hr] or [CM XXXX.X hr]
menu while in the
standby screen.
2.
Hold the
"UP"
and
"DOWN"
button, and then simultaneously press the
"MENU"
button for
one second. Meanwhile,
[TM XXXX.X hr] or [CM XXXX.X hr]
will start blinking and
“CLEAR ”
will
show on the LCD display.
3.
Press
“UP”
or
“DOWN”
. The LCD screen will show
“CLEAR OK”
and the total meter or
compliance meter record will be erased so the value becomes 0.0 hr.
4.
Press
"MENU"
to confirm the reset function.
5.
Press
“START/STANDBY”
button to go back to standby screen or leave the device to
automatically go back to standby screen 5 seconds later.
NOTE –
When the compliance meter has been cleared, the total meter will also be cleared at
the same time, and vice versa.
3. To Download the Compliance Meter and Total Meter Data
Please refer to
Easy Compliance 2.1 CPAP Data Management Software
instruction. Use the
USB cable to connect the device with a computer which has been tested and approved by
IEC60601-1 or IEC60950 standard.
CAUTION
– Do not connect the device to the personal computer for data downloading during
the treatment. This may cause the CPAP system failure.
NOTE –
This additional instruction is separated because of concerns with rental service
management and accidental wrong setting by others. Please tear off this page from
manual before deliver this product to end-user.