Error 001
Error 002
Warn 001
Warn 002
close to mask or flexible tube is
low.
Water Leakage
1. Assembly of water chamber is
not proper or incorrectly.
2. Water chamber or removable
base plate worn out.
1.
Remove the water chamber from the
heated humidifier, pull out the water and
reassemble the water chamber again,
make sure the removable base plate is
right fit into water chamber, and fill the
water till it reach fill line and check if it
still leak or not.
2.
Replace a new water chamber.
Error / Warning Messages show in LCD.
Message type
Definition
Message in LCD
Error for abnormal system settings
Error:
Primary function can’t
execute.
Error for flow generator failure
Out of system memory
Warning:
System memory is nearly full
NOTE:
When the warning message appears, contact your physician or equipment provider to download the
memory data and reset the meter.