EN - 94
SE-7314-2
Turn unit “off” and allow
vacuum to drop.
Disconnect from power source. Disconnect tubing and remove
container from holder.
1
3
2
Carefully remove lid and
empty contents.
Remove filter cartridge and
4-1/2” (11.43 cm) tubing and
set aside.
4
5A
Filter MUST NOT get wet. The
filter material cannot be
removed from the elbow.
CLEANING INSTRUCTIONS
WARNING
To prevent possible risk of infection from contaminated cleaning/disinfection solutions, always prepare fresh solution for each cleaning cycle and discard solution
after each use.
NOTE
– Disinfection information is based on AARC Clinical Practice Guideline Suctioning of the Patient in the Home.
Disposable Collection Container (with Internal Filter Cartridge)
NOTE
– The disposable collection container and lid are meant for single-patient use.
5B
WARNING
- Do not remove
float ball from lid. If
removed, float ball may
pose a choking hazard.
Soak in 1 part vinegar (>=5%
acetic acid concentration) to 3
parts water (131˚F-149ºF/55˚C-
65˚C) solution for 60 minutes.
Rinse with clean, warm water
and air dry.
7
Wash container, lid and splash
guard in warm water/
dishwashing solution. Rinse
with clean, warm water.
6
NOTE
– The disassembled container may also be washed in a dishwasher, top shelf only, using a cycle with a water temperature between 131ºF-149ºF/55ºC-65ºC.
Reusable Collection Container (with External Bacteria Filter)
Turn unit “off” and allow
vacuum to drop.
Disconnect from power source. Disconnect tubing and remove
container from holder.
1
3
Carefully remove lid and
empty contents.
Remove bacteria filter, elbow,
and 4-1/2” (11.43 cm) tubing
and set aside.
Wash jar, lid, o-ring, and
overflow valve in a solution of
warm water with a mild, liquid
detergent (e.g. Dawn or
Palmolive) and rinse with
clean, warm tap water. Then
disinfect using one of the
following methods.
4
5
6
2
For single patient use:
1. Soak in 1 part vinegar (>=5% acetic acid concentration) to 3 parts water (131ºF-149ºF/55ºC-65ºC) solution for 60 minutes. Rinse with clean, warm water and air dry in a
clean environment.
2. Soak with a commercial (bacterial-germicidal) disinfectant. Follow disinfectant manufacturer’s recommended dilution rates and instructions carefully.
For multi patient use:
1. After parts are completely dry, place jar and lid in autoclave with the open end down. Ensure parts are not touching. Run one sterilization steam cycle at 250°F (121°C)
for 15 minutes.
NOTE
–
Jar is guaranteed up to 30 cycles of autoclave sterilization at the indicated conditions.
2. Dispose of and replace filter, tubing and elbow between patients.