Operational Characteristics
12
Make sure that the calibration adapter is correctly seated. To this end, manually
tighten the calibration adapter to a point where the sealing line leaves a permanent
mark on the splash guard.
Nitrogen, synthetic air or fresh air (hydrocarbon content <50 ppm) can be used for
zero point calibration.
Commercially available calibration gas can be used to calibrate the respective gas
category (methane, propane, ethylene). The highest accuracy is achieved using
test gas concentrations of 40 to 70 percent of the measurement span.
The infrared gas transmitter Dräger PIR 3000 can also be used to measure other
gases than mentioned above. For detailed information, refer to "Substitute Gas
Calibration" on page 16.
Select the measured gas in the corresponding table and determine the
corresponding gas category.
Set the gas transmitter to the determined gas category.
Where possible, calibration gas should match with the measured gas for span
calibration. In exceptional cases, span calibration can be carried out using a
suitable substitute gas and the associated calibration factor. The suitable substitute
gas as well as the associated calibration factor is shown in the table "Substitute
Gas Calibration" on page 16.
Select the substitute gas (gas category) and the calibration factor in table
"Substitute Gas Calibration" on page 16.
Multiply the concentration of the substitute gas by the calibration factor to get the
gas concentration to be set.
Example:
Measured gas:
n-octane
Gas category:
Propane (see table "Substitute Gas Calibration",
page 16)
Calibration factor:
1.8 (in table "Substitute Gas Calibration", page 16)
Span gas concentration:
40 %LEL propane (bottle concentration)
Setting:
40 %LEL x 1.8 (calibration factor) = 72 %LEL
CAUTION
Never inhale test gas. Danger to health!
Observe the safety information in the corresponding safety data sheets. Ensure
that gases are vented or otherwise guided outside the building.
Summary of Contents for PIR 3000
Page 1: ...Dräger PIR 3000 en Infrared Gas Transmitter Instructions for Use 00123812_02 eps 0 S ...
Page 25: ...25 Technical Data Dimensions 01123812_01_en eps 0 S aprox 170 mm ø 37 mm M25 x 1 5 or 3 4 NPT ...
Page 28: ...ATEX Approval 28 ATEX Approval ...
Page 29: ...29 ATEX Approval ...
Page 30: ...ATEX Approval 30 ...
Page 31: ...31 ATEX Approval ...
Page 32: ...ATEX Approval 32 ...
Page 33: ...33 ATEX Approval ...
Page 34: ...ATEX Approval 34 ...
Page 35: ...35 ATEX Approval ...
Page 36: ...ATEX Approval 36 ...
Page 37: ...37 ATEX Approval ...
Page 38: ...ATEX Approval 38 ...
Page 39: ...39 ATEX Approval ...
Page 40: ...ATEX Approval 40 ...
Page 41: ...41 ATEX Approval ...
Page 42: ...ATEX Approval 42 ...
Page 43: ...43 ATEX Approval ...
Page 44: ...ATEX Approval 44 ...
Page 45: ...45 ATEX Approval ...
Page 46: ...ATEX Approval 46 ...
Page 47: ...47 ATEX Approval ...
Page 48: ...IECEx Approval 48 IECEx Approval ...
Page 49: ...49 IECEx Approval ...
Page 50: ...IECEx Approval 50 ...
Page 51: ...51 IECEx Approval ...
Page 52: ...IECEx Approval 52 ...
Page 53: ...53 IECEx Approval ...
Page 54: ...UL Approval 54 UL Approval ...
Page 55: ...55 UL Approval ...
Page 56: ...UL Approval 56 ...
Page 57: ...57 CSA Approval CSA Approval ...
Page 58: ...CSA Approval 58 ...
Page 59: ...59 CSA Approval ...
Page 60: ...Declaration of Conformity 60 Declaration of Conformity ...
Page 61: ......
Page 62: ......
Page 63: ......