7
9
Changing dust and water filter
10 Technical Specifications
Environmental Conditions
During operation
–30 to 50
o
C / –22 to 122
o
F
700 to 1300 hPa
10 to 90% relative humidity
Conditions for stor-
age
0 to 40
o
C / 32 to 104
o
F
30 to 80% relative humidity
Ingress protection
IP 65
Pac 3500
Operating times
2 years (typical at 25
o
C)
Pac 5500
Operating times
unlimited
Battery life (typical at
25
o
C)
8 hours of use per day,
1 minute alarm per day:
CO, H
2
S: > 2 years typical
O
2
: > 12 month typical
Intensity of alarm
typical 90 dB (A) at a distance of 30 cm (11.8 in. )
Dimensions
64 x 84 x 20 (battery compartment 25) mm
2.5 x 3.3 x 0.8 (battery compartment 1) in.
Weight
106 g (3.8 oz.)
Approvals
(see “Approvals” on page 171)
!
00223806_04.eps
11 Sensor Specifications
_____________
Date of production: see label at the back case.
12 Accessories
CO
H
2
S
O
2
Reproducibility
Zero point:
d
±2 ppm
d
±0.1 ppm
d
±0.2 vol.-%
Sensitivity:
d
±2 % of
measured value
d
±5 % of
measured value
d
±1 % of
measured value
Long-term drift (20
o
C)
Zero point:
d
±2 ppm/a
d
±0.2 ppm/a
d
±0.5
vol.-%/a
Sensitivity:
d
±1 % of
measured value /
month
d
±1 % of
measured value /
month
d
±1 % of
measured value /
month
Please be aware of possible sensor cross sensitivities (see Data Sheet CO:
9023816, H
2
S: 9023819, O
2
: 9023820).
Description
Order-code
Connecting Cradle, complete with USB cable and Pac Vision
software
83 18 587
Calibration Adapter
83 18 588
Lithium battery
45 43 808
Dust and water filter
45 43 836
Leather carrying case
45 43 822
Bump Test Station, complete with test gas cylinder 58 L
(gas type by customer request)
83 18 586
E-Cal device module for connection of 4 Dräger Pac 1000 to
7000 to a E-Cal Master Station or to Module Adapter
83 18 589