Maintenance intervals

Intervaly údržby

Testing the system

Test mode facilitates installation of the module, enables 
function tests to be performed at any time on the 
individual relay channels and permits manual adjustment 
of the output states. Test mode is only available when the 
module is active and not in the event of a power failure or 
if the module is defective.

After a reset, the module always enters normal mode. 
The Test key must be pressed in order to switch over to 
test mode.
Test mode is indicated by a red LED (to the left next to the 
Test key). In this mode, the relevant output status can be 
permanently switched over with the keys until test mode 
is deactivated.
Output changes received via the LON

®

 network only take 

effect after quitting test mode.
Test mode is terminated by pressing the Test key again. 
All outputs are updated in accordance with the received 
message.
If a relay key is pressed in while in normal mode, the 
output status of the corresponding relay is switched over 
while the key is pressed.
Test mode must be deactivated before ending work on 
the network.

Maintenance intervals

Regularly, at the latest once every six months:

Operating check and visual inspection by specially 
trained experts. We recommend that you take out a 
Service Contract with Dräger Service.

00

637296   

   

 

19

1

2

3

5

6

7

8

9 10 11 12 13 14 15 16 17 18

20 21 22

24 25 26 27 28 29

31 32 33 34 35 36

1

Test

Service

Digitaleingang

2

3

4

5

6

7 8

Zkouška systému

Zkušební režim usnad

ň

uje montáž modulu, umož

ň

uje 

provád

ě

ní funk

č

ních zkoušek jednotlivých reléových 

kanál

ů

 v kteroukoli dobu a dovoluje ru

č

ní nastavení stav

ů

 

výstup

ů

. Zkušební režim je možný pouze, je-li modul 

aktivní; nikoli p

ř

i výpadku napájení nebo závad

ě

 modulu.

Po resetování se modul nachází vždy v b

ě

žném režimu. 

P

ř

epnutí do zkušebního režimu se provádí tla

č

ítkem Test.

Zkušební režim je signalizován 

č

ervenou kontrolkou 

(vlevo od tla

č

ítka Test). V tomto režimu lze tla

č

ítky 

natrvalo p

ř

epínat ten 

č

i onen stav výstupu, dokud 

zkušební režim op

ě

t nebude deaktivován.

Zm

ě

ny výstup

ů

 p

ř

ijaté prost

ř

ednictvím sít

ě

 LON

®

 jsou 

ú

č

inné až po opušt

ě

ní zkušebního režimu.

Zkušební režim se ukon

č

í op

ě

tovným stisknutím tla

č

ítka 

Test. Všechny výstupy jsou aktualizovány podle p

ř

ijatého 

hlášení.
P

ř

i stisknutí tla

č

ítka n

ě

kterého relé v b

ě

žném režimu je na 

dobu stisknutí tla

č

ítka p

ř

epnut stav výstupu p

ř

íslušného 

relé.
Zkušební režim musí být p

ř

ed ukon

č

ením práce v síti 

deaktivován.

Intervaly údržby

Pravideln

ě

, nejmén

ě

 jednou za p

ů

l roku:

Funk

č

ní a vizuální kontrola celého za

ř

ízení odborníky. 

Doporu

č

ujeme vám uzav

ř

ení servisní smlouvy se 

servisem DrägerService.

Summary of Contents for G 50770

Page 1: ...00137296 Installation Instructions Instructions for Use Mont n n vod N vod k pou it Gas Management Monitor R...

Page 2: ...ystem 10 Maintenance intervals 10 Technical data 11 Acceptance and handover to the operator 12 System overview 13 Obsah Pro va i bezpe nost 3 P edpokl dan pou it 4 Mont 4 Mont Monitoru R 4 P prava p i...

Page 3: ...endations given above The warranty and liability provisions of the terms of sale and delivery of Dr gerwerk Aktiengesellschaft are likewise not modified by the recommendations given above Dr gerwerk A...

Page 4: ...L x H 160 x 360 x 100 mm Monitor R dimensions W x L x H 86 x 105 x 58 mm 00137296 00237296 105 58 86 19 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36...

Page 5: ...cable types should be agreed with the design department for each project LON Local Operating Network regi stered trade mark of the Echelon com pany P prava p ipojovac ho kabelu Provozn nap t Voliteln...

Page 6: ...hen using 24V DC Ensure correct polarity 00337296 Ub Ub Schirm1 Shield 1 La1 Lb1 00437296 IN OUT 1 2 3 4 LA LB S LA LB S LA LB S LA LB S LA LB S F1 F 6 3 A 24V Network 24V Network Syst mov kabel s t G...

Page 7: ...al 23 Ground to connection 24 NC NC 1 Test Service Digitaleingang 2 3 4 5 6 7 8 13 23 33 43 34 44 14 24 Relais 53 63 73 83 74 84 54 64 Relais Bus SH Bus B Bus A OUT IN 24V 24V NC neobsazeno P pojky Vs...

Page 8: ...ease note that for fault free operation in the network a terminating resistor LON terminator G 50 769 or Monitor CC G 50 718 is required Ulo te kabel v krabici tak aby se elektronick sou sti a st n n...

Page 9: ...s status with the appropriate configuration can be transmitted via a relay signal fault relay In this case relay No 8 is used Provoz kolem rel ov ho modulu je odes lat jednotliv a hromadn hl en s t ze...

Page 10: ...ck and visual inspection by specially trained experts We recommend that you take out a Service Contract with Dr ger Service 00637296 19 1 2 3 5 6 7 8 9 10 11 12 13 14 15 16 17 18 20 21 22 24 25 26 27...

Page 11: ...ing current DC1 24V 150 mA Max switching current DC13 24V 50 mA Code UMDNS Universal Medical Device Nomenclature System 18 046 Technick daje Po adavky na okoln prost ed Teplota 15 o C a 40 o C Relativ...

Page 12: ...st be recorded in writing Following acceptance the operational system complete with associated documents is handed over to the operator Handover must be recorded in an official document Schv len a p e...

Page 13: ...W UP Set 310 G 50 742 HW UP Set 400 G 50 743 AP Set 250 G 50 745 AP Set 310 G 50 746 AP Set 400 G 50 747 AP Housing 360 G 50 748 P ehled syst mu Pro datovou s t syst mu zen plyn se dod vaj tyto sou st...

Page 14: ...rail G 50 731 Communication cable e g GM network system cable G 50 750 Designation description Part No Sou sti syst mu jsou mezi sebou propojeny s t LON V jednom segmentu sm b t vz jemn propojeno max...

Page 15: ...15 This page intentionally left blank T to strana bola ponechan z merne pr zdn...

Page 16: ...beck 49 451 8 82 0 FAX 49 451 8 82 20 80 http www draeger com 90 55 461 MA GA 6918 020 en cs Dr gerwerk AG Co KGaA Edition 1 2015 10 Subject to alteration 90 55 461 MA GA 6918 020 en cs Dr gerwerk AG...

Reviews: