Operation
60
Instructions for use D-Vapor
/
D-Vapor 3000
7
O
2
monitor switched on. Alarm limits set.
8
When CO
2
accessory equipment (CO
2
module)
is used for dosing CO
2
on the anesthesia
workstation, CO
2
may be conducted through by
the vaporizer. Observe instructions for use of
the anesthesia workstation.
Checking for damage
Check the vaporizer for visible damage.
Check the power cable and mains plug for
visible damage.
Check that the power cable is not trapped
between the vaporizer and the anesthesia
workstation. Observe the instructions for use for
the anesthesia machine.
Checking the filling level
Filling level in sight glass sufficiently high above
minimum mark – must not exceed maximum
mark.
Checking the filling device
Filling device for the Baxter SAFE-FIL filling
system for desflurane (Suprane): Sealing plug
inserted and engaged.
Filling device for the Piramal Fill filling system
for desflurane (Torrane): Opening on the filling
device closed. Screw cap correctly fitted and
tightened.
Checking the connection system
Plug-in connector:
Plug-in adapter sits evenly on the seals.
Locking lever swung to the left. Vaporizer is
hanging vertically on machine, when viewed
from front and side, and cannot be lifted off.
Other connectors:
Vaporizer connected firmly and securely on
anesthesia workstation. Vaporizer is hanging or
standing upright and is secured against tilting or
falling.
Direction of flow corresponds to arrow on
vaporizer.
If there are several connectors, check
connector allocation and Interlock
1
All connectors allocated.
2
The following unallocated connectors must not
be open during operation:
WARNING
Risk of patient injury
Incorrect output concentrations will not be de-
tected when operating from the external fresh-
gas outlet.
An anesthetic gas monitor must be used if the
vaporizer is used with anesthesia machine via
the external fresh-gas outlet.
WARNING
In accordance with ISO 21647 or
ISO 80601-2-55, use of a continually measur-
ing oxygen monitor with an alarm system for
detecting insufficient oxygen supply, e.g., due
to leakage, is required.
D-Vapor
D-Vapor 3000
Permanent connections –
ISO conical adapter
–
Plug-in connectors
without valve function
Plug-in connectors
without valve function
WARNING
Risk of insufficient supply to the patient
Non-allocated, open connections will result in
a loss of fresh gas and anesthetic agent. The
supply to the patient will be interrupted.
Non-allocated connectors must not be open
during operation.
Summary of Contents for D-Vapor
Page 14: ...14 Instructions for use D Vapor D Vapor 3000 This page has been left blank intentionally ...
Page 15: ...Instructions for use D Vapor D Vapor 3000 15 Use Use Intended use 16 Environment of use 16 ...
Page 26: ...26 Instructions for use D Vapor D Vapor 3000 This page has been left blank intentionally ...
Page 76: ...76 Instructions for use D Vapor D Vapor 3000 This page has been left blank intentionally ...
Page 88: ...88 Instructions for use D Vapor D Vapor 3000 This page has been left blank intentionally ...
Page 114: ...114 Instructions for use D Vapor D Vapor 3000 This page has been left blank intentionally ...
Page 116: ...116 Instructions for use D Vapor D Vapor 3000 This page has been left blank intentionally ...
Page 119: ...Instructions for use D Vapor D Vapor 3000 119 This page has been left blank intentionally ...