Instructions for use D-Vapor/D-Vapor 3000
49
Preparation
Several Selectatec-compatible plug-in
connectors
1
Switch off vaporizers on other plug-in
connectors. Set the control dial to the
0
position.
Direct contact on the side Interlock pins is required
for the Interlock system to operate. For triple plug-
in connectors with built-in detection of the locking
status between the outer plug-in positions, the
middle plug-in position may remain unoccupied.
Several Dräger Auto Exclusion plug-in
connectors
A D-Vapor or D-Vapor 3000 with Auto Exclusion
plug-in adapter can be connected to any free
Dräger Auto Exclusion plug-in position.
Positioning the vaporizer on the plug-in
connector
1
Hold the vaporizer in the vertical position with
both hands and lower it gently onto the pins on
the plug-in connector.
If loss of fresh gas, leaks, excessively low
concentrations of anesthetic agent or jamming of
the Interlock locking device occur, remove the
vaporizer, see page 67, check the lever position
and the engaging mechanism, then reconnect the
vaporizer.
2
Swing the locking lever (B) clockwise by 90°
until it engages. The vaporizer is then secured
and cannot be removed.
3
Press the
0
button and set the control dial to the
0
(C) position.
WARNING
Risk of patient injury
A malfunction in the Interlock system can en-
danger the patient by overdosing or mixing
anesthetic agents.
On the Interlock S the side pins, which are ex-
tended when a vaporizer is switched on, must
align axially. These pins must be next to each
another on the vaporizers, and under no cir-
cumstances above each other.
059
0
CAUTION
Risk of injury by trapping fingers
Take care when lowering the vaporizer onto its
plug-in connector.
WARNING
Risk of patient injury
Risk of fresh gas and anesthetic vapor escap-
ing.
If the vaporizers are not correctly positioned
or not correctly locked, this can result in a
loss of fresh gas due to leaks, in low concen-
trations or in jamming of the Interlock inter-
lock device.
The plug-in adapter must sit evenly and hori-
zontally on the
sealing rings.
028
0
2
3
4
B
C
Summary of Contents for D-Vapor
Page 14: ...14 Instructions for use D Vapor D Vapor 3000 This page has been left blank intentionally ...
Page 15: ...Instructions for use D Vapor D Vapor 3000 15 Use Use Intended use 16 Environment of use 16 ...
Page 26: ...26 Instructions for use D Vapor D Vapor 3000 This page has been left blank intentionally ...
Page 76: ...76 Instructions for use D Vapor D Vapor 3000 This page has been left blank intentionally ...
Page 88: ...88 Instructions for use D Vapor D Vapor 3000 This page has been left blank intentionally ...
Page 114: ...114 Instructions for use D Vapor D Vapor 3000 This page has been left blank intentionally ...
Page 116: ...116 Instructions for use D Vapor D Vapor 3000 This page has been left blank intentionally ...
Page 119: ...Instructions for use D Vapor D Vapor 3000 119 This page has been left blank intentionally ...