Preparation
46
Instructions for use D-Vapor/D-Vapor 3000
Vaporizer with Piramal Fill filling system for desflurane (Torrane)
Starting the filling procedure
Unscrew the cap from the anesthetic agent
bottle. Check that the O-ring on the bottle
connection is not damaged.
1
Remove the screw cap (A) from the filling port
by unscrewing.
2
Insert the anesthetic agent bottle (B) firmly into
the filling port, ensuring that the threads
engage. Turn the bottle clockwise until a spring
resistance can be felt. Do not tilt the bottle while
turning it.
3
Turn the anesthetic agent bottle clockwise
against the noticeable spring resistance until a
higher noticeable resistance is reached.
CAUTION
Vaporizer not fillable
Only use desflurane bottles from Piramal. These
bottles have the appropriate filling connector. Oth-
erwise, the Vaporizer will not be filled.
Do not use excessive force.
WARNING
Danger to the user
The valve pin in the filling device must only be
actuated by the Piramal desflurane bottle. Do
not use any other object to depress the valve
pin. Otherwise, desflurane may spurt.
WARNING
Spraying of anesthetic agent
If the bottle is not properly inserted into the
filling port, the thread may jam and the bottle
will open while the vaporizer is still closed. In
this case, the anesthetic agent may spray out
with considerable pressure.
Make sure that the filling connector of the bot-
tle is correctly inserted into the filling port of
the vaporizer and that the bottle threads are
properly engaged with the port threads before
screwing it into the filling system.
079
0
Vol.%
A
080
081
WARNING
Spraying of anesthetic agent
If the bottle continues to be turned after the
bottle valve has been opened, anesthetic
agent may spurt out with considerable force.
Do not continue to turn.
0
Vol.%
B
0
Vol.%
Summary of Contents for D-Vapor
Page 14: ...14 Instructions for use D Vapor D Vapor 3000 This page has been left blank intentionally ...
Page 15: ...Instructions for use D Vapor D Vapor 3000 15 Use Use Intended use 16 Environment of use 16 ...
Page 26: ...26 Instructions for use D Vapor D Vapor 3000 This page has been left blank intentionally ...
Page 76: ...76 Instructions for use D Vapor D Vapor 3000 This page has been left blank intentionally ...
Page 88: ...88 Instructions for use D Vapor D Vapor 3000 This page has been left blank intentionally ...
Page 114: ...114 Instructions for use D Vapor D Vapor 3000 This page has been left blank intentionally ...
Page 116: ...116 Instructions for use D Vapor D Vapor 3000 This page has been left blank intentionally ...
Page 119: ...Instructions for use D Vapor D Vapor 3000 119 This page has been left blank intentionally ...