![Dräger D-Vapor Instructions For Use Manual Download Page 112](http://html.mh-extra.com/html/drager/d-vapor/d-vapor_instructions-for-use-manual_2526743112.webp)
Description
112
Instructions for use D-Vapor/D-Vapor 3000
Correction of the flow measurement under the influence of desflurane
Hot wire volume monitors, which are not
compensated for desflurane, measure volumes
that are too great when the desflurane
concentration is high. See instructions for use for
the monitor used.
Correcting values for the minute volume MV
Correcting values for tidal volume VT
Example:
– Desflurane concentration 8 Vol%
– Displayed minute volume MV 10 L/min
MV
corrected
= Displayed MV x
100 – 2 x desflu-
rane concentration
100
VT
corrected
= Displayed VT x
100 – 2 x desflu-
rane concentration
100
MV
corrected
=
10
x
100 – 2 x 8
100
MV
corrected
=
10 x 0.84
=
8.4 L/min
Summary of Contents for D-Vapor
Page 14: ...14 Instructions for use D Vapor D Vapor 3000 This page has been left blank intentionally ...
Page 15: ...Instructions for use D Vapor D Vapor 3000 15 Use Use Intended use 16 Environment of use 16 ...
Page 26: ...26 Instructions for use D Vapor D Vapor 3000 This page has been left blank intentionally ...
Page 76: ...76 Instructions for use D Vapor D Vapor 3000 This page has been left blank intentionally ...
Page 88: ...88 Instructions for use D Vapor D Vapor 3000 This page has been left blank intentionally ...
Page 114: ...114 Instructions for use D Vapor D Vapor 3000 This page has been left blank intentionally ...
Page 116: ...116 Instructions for use D Vapor D Vapor 3000 This page has been left blank intentionally ...
Page 119: ...Instructions for use D Vapor D Vapor 3000 119 This page has been left blank intentionally ...