93
Error List
Babylog 8000
Display error
number
Monitoring PCB
error number
Meaning
505
CPU PCB software error
506 to 513
CPU PCB software error.
800
FiO
2
too high by more than 10
vol.%.
801
FiO
2
too low by more than 10
vol.%.
802
FiO
2
measurement deviation
between master (CPU PCB and
I/O PCB) and Monitoring PCB is
too high.
803
Measured Pinsp to potentiometer
setting Pinsp too high by more
than 25 mbar (cmH
2
O).
804
Pinsp greater than 110 mbar
(cmH2O).
805
Pinsp less than -5 mbar (cmH2O)
806
Flow measurement on the Flow
PCB is faulty
807
Potentiometer evaluation is faulty:
Master (CPU PCB and
Frontcontroller PCB) and
Monitoring PCB read different
values.
809
Battery or battery charge faulty
(battery voltage < 7 V)
Table 1: Device error messages