Calibration / LED / Potentiometer Test
•
Assemble the Babylog 8000/Babylog 8000 plus/Babylog 8000 SC including the
tubing system but without the humidifier.
•
Switch the device on.
•
Note the software version and the operating hours down.
Communication conversion kit y/n
(is displayed under the operating hour indication
after power-on).
Important
: 8306065 or 8411108 (see nameplate option field) without central-alarm
function. The sockets for the communication interface may be present even if this
function is not available.
•
O
2
calibration.
•
Flow sensor calibration (only in the Babylog 8000).
LED and horn test (press and hold the “Reset Check” : after 2 s, each LED and the
stand-by horn are triggered.).
Important
:
Alarm LEDs = 1 yellow and 2 red LEDs
Trigger LEDs = 3 yellow LEDs
Display backlight.
•
Testing the keys (check respective reaction of the equipment).
Potentiometers (no fault message about potentiometer malfunction).