Testing the Mixer
•
Carry out an O
2
calibration.
•
Test set-up (see Figure below):
Fig. 9:
Mixer
•
Select the following inspiratory flows from the “list” menu: 1.2 / 1.3 / 1.4 / 1.5 / 1.6
/ 1.7 / 1.8 / 1.9 / 2.0 / 4 / 5 / 6 / 8 / 9 / 9.9 / 11 / 12 / 13 / 14 / 15 / 30
Allow approx. 1 min between setting and measurement of flow values (< 2 L/min). If
no values is set, the alarm “FiO
2
deviation” is allowed to occur.
15
Testing the Patient System Heating
•
Put your hand against the expiratory valve block.
The temperature should exceed the room temperature.