Replacing Non-Repairable Items
•
Set the pressure reducer to 1.7
-0.1
bar.
•
Set the insp. flow adjuster to 30 L/min. The manometer should indicate at least 1.3 bar.
•
Lock both adjusting rings by pressing them in direction of the arrows
C
.
•
Make sure that the hose system is free of leaks.
Fig. 21:
Locking the Pressure Reducer
•
Re-assemble the Babylog 8000/Babylog 8000 plus/Babylog 8000 SC and it return it to service ready-
for-use.