Leak Test
•
Adjust the following parameters on the Babylog 8000/
Babylog 8000 plus/Babylog 8000 SC (see Figure below):
Fig. 6:
Leak Test
•
Select
pressure
from the menu “pressure curve” and “measured values”.
•
Switch off occurring alarms and warnings by pressing the “Reset/Check” key.
•
Press and hold the
manual inspiration
key.
A pressure is now built up for 5 s. During this time, a pressure of at least 75 mbar
must be achieved.
60
1
2
1
80
0