Replacing Non-Repairable Items
Babylog 8000
Fig. 19:
Air-Supply Test Set-Up
•
Carry out the measurement with a manometer and a secretion sightglass at the pneumatics outlet.
•
Slowly turn the adjusting ring of the air pressure reducer clockwise. The message “no compressed
air” on the LCD disappears.
•
Set the pressure reducer to 1.7
-0.1
bar.
•
Set the insp. flow adjuster to 30 L/min. The manometer should indicate at least 1.3 bar.
•
Disconnect the Babylog 8000/Babylog 8000 plus/Babylog 8000 SC from the air supply and connect it
to the oxygen supply.
•
Adjust the following parameters:
Mode = CPAP; PEEP/CPAP adjuster = 0 mbar; insp. flow = 1 L/min.
Fig. 20:
O
2
-Supply Test Set-Up
•
Carry out the measurement with a manometer and a secretion sightglass at the pneumatics outlet.
•
Slowly turn the adjusting ring of the O
2
pressure reducer clockwise. The message “no oxygen” on the
LCD disappears.