Supplement
Fabius plus SW 3.n
43
Technical data
Vaporizer interface
The data for the PAW, MVe, and VTe parameters in the
following table have changed.
Measured value or
waveform
Range
Resolution Accuracy
Condition
PAW
Airway pressure
(numeric)
-20 to 99 cmH
2
O
(hPa)
1 cmH
2
O
(hPa)
±4 %
1)
1) Max. ±4 % of the measured value or ±2 cmH
2
O (±2 hPa), depending on which value is higher.
Airway pressure
(waveform)
0 to 99 cmH
2
O
(hPa)
Pressure gauge
(mechanical)
-20 to 80 cmH
2
O
(hPa)
5 cmH
2
O
(hPa)
±(2 % of scale value +
4 % of displayed
value)
MV
e
Expiratory minute
volume
1 to 32.0 L/min
0.1 L/min
±15 % or ±0.2 L/min,
depending on which
value is higher
<1 L/min
0.1 L/min
±(20 mL x respiratory
rate) or ±0.2 L/min,
depending on which
value is higher
VT
e
Expiratory tidal
volume
100 to 1500 mL
1 mL
±15 %
<100 mL
1 mL
±20 mL
Note: If the endtidal desflurane concentration increases to above 12 %, the measurement accuracy of
the tidal volume and minute volume can deviate by more than 15%.
Freq
Respiratory rate
2 to 99 bpm
(1/min)
±1 bpm
(1/min)
±1 bpm (±1 1/min) of
the setting or ±5%,
depending on which
value is higher
FiO
2
O
2
measurement
in the mainstream
10 to 100 Vol%
1 Vol%
±2.5 Vol% +2.5 % of
the measured values
in accordance with
ISO 21647 and
ISO 80601-2-55
Based on the
ambient pressure
during calibration