Operation
22
Supplement
Fabius plus SW 3.n
Operation with non-rebreathing system
Diverting the fresh-gas flow to the non-rebreathing
system:
1
Set the switch to
.
The switch points in the direction of the non-
rebreathing system.
2
Set the fresh-gas flow.
To prevent rebreathing, the fresh-gas supply
must be at least double the minute volume.
3
Operate the non-rebreathing system according
to the corresponding instructions for use.
Operation with the compact breathing
system (COSY)
Before using a controlled ventilation mode
(
Volume Control
,
Pressure Control
,
Pressure
Support
,
SIMV/PS
) or the
ManSpont
ventilation
mode:
1
Set the switch to
COSY
.
The switch points in the direction of the fresh-
gas inlet.
2
Screw the sample line back to the Y-piece on
the breathing circuit.
When the switch of the external fresh-gas outlet is
set to operation with a non-rebreathing system
, the ventilator lock prevents a controlled
ventilation mode (
Volume Control
,
Pressure
Control
,
Pressure Support
,
SIMV/PS
) from being
started.
The following alarms are issued:
From software version 3.37:
–
VENTILATOR FAIL !!!
–
CHECK APL / EXT. FGO !!!
With older software versions:
–
VENTILATOR FAIL !!!
–
CHECK APL VALVE !!!
Ending operation
1
Close all flow control valves on the device.
41718
41673
Pmax=
125 hPa
COSY
CAUTION
Risk of undetected fresh-gas deficiency
When the switch of the external fresh-gas outlet
is set to operation with a non-rebreathing system
[
], no alarm occurs at the start of the
ManSpont
ventilation mode.
Check that the switch of the external fresh-gas
outlet has been set to
COSY
.