Supplement
Fabius plus SW 3.n
25
Configuration
Configuration
Configuration in Standby mode
Leakage test
The following text has changed:
Safety valve test
This test checks the functional integrity of the
safety valves.
The test results are displayed on the screen with
the leakage test results (B).
Troubleshooting in the event of a failed safety
valve test
The following safety information has changed:
Repeat leakage test. Do not operate the device
if the safety valve test continues to fail. Contact
DrägerService or the authorized local service
partner
Results of the compliance test
The value for the system compliance has changed:
WARNING
Risk of unexpectedly occurring overpressure
or negative pressure
A soiled or non-functioning safety valve is
not able to compensate for suddenly
occurring overpressure or negative pressure
in the breathing system.
Perform the leakage test before start-up of
the device. Observe the result of the safety
valve test.
System compliance
[mL/cmH
2
O]
Displayed result
[mL/cmH
2
O]
≤
6.0
Measured value and
PASSED