Assembly and preparation
16
Supplement
Fabius plus SW 3.n
Assembling the breathing system
The following chapters have changed:
Mounting the COSY heater and the
power supply unit (optional)
The information for the assembly instructions has changed:
Follow the assembly instructions (8605899) for the
breathing system heating.
Connecting the breathing hoses and the
filters
The data for the tidal volumes for adults and pediatric patients
have been changed:
1
Select suitable accessories for the respective
patient category.
Adults
Pediatric patients Neonates
Tidal volume
>700 mL
301 to 700 mL
50 to 300 mL
<50 mL
Breathing bag
3 L
2 L
1 L
0.5 L
Breathing circuit
Adults
Pediatric
Neonates
(or pediatric)
Filter
Filter, HMEF, or HME
Use filters with low
resistance and