Troubleshooting
30
Supplement
Fabius plus SW 3.n
The following alarm has been added:
(28)
VENTILATOR FAIL !!!
Fault in APL bypass valve.
Check membrane of the
ventilator and close cover.
Check the connection of
the APL bypass valve for
correct connection and for
leakage. Select
Standby
mode and then switch to
the previous ventilation
mode. Check APL valve
setting.
Switch of the external fresh-
gas outlet with switch and
ventilator lock in the wrong
position (
), instead of
the
COSY
position.
Set the switch of the exter-
nal fresh-gas outlet to the
COSY
position.
Alarm
priority
Alarm
Cause
Remedy
Alarm
priority
Alarm
Cause
Remedy
(26)
CHECK APL / EXT. FGO !!!
Fault in APL bypass valve.
Check membrane of the
ventilator and close cover.
Check the connection of
the APL bypass valve for
correct connection and for
leakage. Select
Standby
mode and then switch to
the previous ventilation
mode. Check APL valve
setting.
Switch of the external fresh-
gas outlet in the wrong posi-
tion (
), instead of the
COSY
position.
Set the switch of the exter-
nal fresh-gas outlet to the
COSY
position.