![Dräger 8413417 Instructions For Use Manual Download Page 36](http://html.mh-extra.com/html/drager/8413417/8413417_instructions-for-use-manual_2526715036.webp)
English / US English
36
Instructions for Use Air Compressor
Recommended separation distances to
portable and mobile RF
telecommunication devices
The separation distances below are in accordance
with IEC 60601-1-2.
Reduced separation distances to
portable and mobile RF
telecommunication devices
The separation distances below are based on tests
performed by Dräger to determine the minimum
separation distances. These reduced separation
distances are valid only for mobile RF
telecommunication devices using the standards
listed.
Max. P
EIRP
(W) 150 kHz to 2.5 GHz
All other frequencies Examples
0.03
0.32 m (1.05 ft)
0.96 m (3.15 ft)
e.g., WLAN 5250 / 5775 (Europe)
0.10
0.58 m (1.90 ft)
1.75 m (5.74 ft)
e.g., WLAN 2440 (Europe)
0.17
0.76 m (2.49 ft)
2.28 m (7.48 ft)
e.g., Bluetooth, RFID 2.5 GHz
0.20
0.82 m (2.69 ft)
2.47 m (8.10 ft)
e.g., WLAN 5250 (not in Europe)
0.25
0.92 m (3.02 ft)
2.76 m (9.06 ft)
e.g., UMTS mobiles
0.41
1.18 m (3.87 ft)
3.53 m (11.58 ft)
e.g., DECT devices
0.82
1.67 m (5.48 ft)
5.00 m (16.40 ft)
e.g., RFID 13.56 MHz
1.00
1.84 m (6.04 ft)
5.52 m (18.11 ft)
e.g., WLAN 5600 (not in Europe)
1.64
2.36 m (7.74 ft)
7.07 m (23.20 ft)
e.g., GSM 1800 / GSM 1900
3.28
3.33 m (10.93 ft)
10.00 m (32.81 ft)
e.g., GSM 900 mobiles, RFID
868 MHz
Mobile communication device using ...
Separation distance
GSM 850, GSM 900, RFID 868 MHz (limited to 2 W ERP)
0.07 m (0.23 ft)
GSM 1800, GSM 1900 (limited to 1 W ERP)
0.05 m (0.16 ft)
UMTS, DECT (limited to 0.25 W ERP)
0.02 m (0.07 ft)
Bluetooth, WLAN 2450, RFID 2450 (limited to 0.1 W ERP)
0.03 m (0.10 ft)