4
Gas Supply Connections
4-4
Narkomed GS Setup and Installation Manual
Draeger Medical, Inc.
Connecting
the Pipelines
Pipeline connections for oxygen, nitrous oxide and an optional third gas
are located on the side of the flowmeter housing. The pipeline inlets are
marked with color-coded labels for each gas. (See the “Gas System Color
Coding” table at the beginning of this section.)
WARNING:
Oil and grease may combine explosively with oxygen or
nitrous oxide. For this reason, oil and grease must never come
in contact with pipelines, cylinders, cylinder valves, gauges,
fittings, etc., which conduct oxygen or nitrous oxide within the
machine. For further information regarding safety
precautions in the use of medical gases, consult Compressed
Gas Association pamphlet P-2 and appropriate sections of
National Fire Protection Association Standard 99.
YOKE
ASSEMBLY
YOKE HANDLE
BOLT
YOKE
HANDLE
YOKE PLUG
CYLINDER
WRENCH
INDEX
PINS
GAS
INLET
CYLINDER
VALVE
SEALING
WASHER
CYLINDER
HEAD
GAS CYLINDER
(COLOR CODED)
SU20035A
Summary of Contents for Narkomed GS
Page 2: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 10: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 40: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 46: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 50: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 52: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 90: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 91: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...