6-28
NMGS
PMC PROCEDURE (continued)
6.15.18
Adjust the Inspiratory Flow until a peak pressure of 80 cm H
2
O is
achieved.
6.15.19
If applicable, set the Pressure Limit Control to within the 30 range.
Readjust within the band as necessary to achieve proper value. Is the
peak pressure at the 30 range within 27 to 33 cm H
2
O? ___ (Y)
6.15.20
Set the Pressure Limit Control to the MIN position. Is the peak pressure
at the MIN range 0 to 15 cm H
2
O? ___ (Y) Return the Pressure Limit
control to MAX.
6.15.21
Loosen the expiratory valve dome, or if equipped with an ultrasonic flow
sensor, open the Y-piece and disconnect the hose attached to the
exhalation valve and blow into it.
6.15.22
Does the Reverse Flow message appear on the display? ___ (Y)
6.15.23
Tighten the expiratory valve dome if applicable, or if equipped with an
ultrasonic flow sensor reconnect the hose between the expiratory valve
and the flow sensor.
6.15.24
Insert a test minute volumeter in between absorber dome and Spiromed,
or exhalation valve and absorber dome if equipped with an ultrasonic
flow sensor.
6.15.25
Open the Y-piece.
6.15.26
Turn the ventilator off and press the ALL STBY key to clear alarms.
6.15.27
Turn the ventilator on and start a stop watch.
6.15.28
Does APNEA-VOLUME appear as a Caution within 13 to 17 seconds?
___ (Y)
6.15.29
Attach a 3 liter breathing bag to the Y-piece.
NOTE:
Bag should be placed on a flat horizontal surface to reduce
artifact volume.
6.15.30
Press the O
2
Flush momentarily to inflate the bellows.
6.15.31
Set the Inspiratory Flow to the MED and readjust as necessary to fully
collapse the bellows.
6.15.32
Observe the operation of each unidirectional valve disc at eye level. Does
the inspiratory valve disc raise only during the inspiration phase, and
the expiratory valve raise only during the exhalation phase? ___(Y)
6.15.33
Is the tidal volume on the volume monitor and on the test volumeter
within 20 % of each other? ___ (Y)
6.15.34
Does the volume monitor display 10 BPM? ___ (Y)
6.15.35
Does the display correctly track the Breathing Pressure waveform?
___ (Y)
6.16 BELLOWS ADULT - If applicable
6.16.1
Is the tidal volume indicated on the test volumeter 960 to1440 ml? ___(Y)
(
9
)
(
9
)
(
9
)
(
9
)
(
9
)
(
9
)
(
9
)
Summary of Contents for Narkomed GS
Page 2: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 10: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 40: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 46: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 50: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 52: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 90: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 91: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...