6-16
NMGS
PMC PROCEDURE (continued)
6.6.13
Are there two (2) spring clips on the absorber rods? ___ (Y)
6.6.14
Inspect the following: canisters, canister gaskets, dust cup and O-ring,
and soda lime.
6.6.15
Are the canisters, canister gaskets, dust cup and O-ring, and soda lime
in good condition? ___ (Y)
6.6.16
Verify the cm H
2
O gauge at zero (0) and readjust if necessary.
NOTE:
The small slotted screw is the zero adjust.
6.6.17
Reinstall the O
2
sensor plug into the inspiratory valve dome adapter.
6.6.18
Remove the 15-mm connector from the FRESHGAS OUTLET.
6.6.19
Is the Freshgas Outlet assembly in good condition? ___ (Y)
6.6.20
Reconnect the 15-mm connector to the FRESHGAS OUTLET.
6.7
HIGH PRESSURE LEAK
NOTE:
Minimum cylinder pressures required for High Pressure Leak tests are:
O
2
, Air, O
2
-HE, N2, HE: 1000 psi / N
2
O, CO
2
: 600 psi;
6.7.1
Turn the machine main switch to Standby.
6.7.2
Verify the Auxiliary Oxygen flow control valve is closed.
6.7.3
Disconnect all pipeline supply hoses at the wall outlets.
6.7.4
Open then close and remove each cylinder and if applicable remove the
yoke plug from each additional yoke assembly.
6.7.5
Note the reading on each the cylinder pressure gauge and start a stop
watch.
6.7.6
Are the two (2) yoke index pins installed securely in each yoke? ___(Y)
6.7.7
Is the proper gas I.D. label affixed to each yoke? ___ (Y)
6.7.8
After two (2) minutes, is the pressure loss for each gas equal or less than
50 psi? ___(Y)
6.7.9
Verify the presence of only one (1) cylinder washer, then reattach and
secure the cylinders to each yoke assembly, then open each cylinder
valve.
6.8
BREATHING SYSTEM
6.8.1
Breathing System Leak/Exclusion
6.8.1.1
Close all flow control valves.
6.8.1.2
Set the AUTO/MAN selector to BAG.
6.8.1.3
Close the APL valve.
6.8.1.4
Interconnect a 22 mm hose (P/N 9995132) between the
inspiratory valve and expiratory valve or expiratory port on
the ultrasonic flow sensor, if applicable.
(
9
)
Summary of Contents for Narkomed GS
Page 2: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 10: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 40: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 46: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 50: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 52: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 90: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 91: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...