NMGS
PMC PROCEDURE (continued)
6-33
6.19.2.12
What is the pressure on the test gauge? ___ cm H
2
O (0 to -0.5)
6.19.2.13
Remove the test equipment, re-cap the scavenger adapter port
or remove the scavenger adapter and reconnect the scavenger
hose.
6.20 A/C SCAVENGER - 6 Month Service Interval; Due Date ______, If applicable
NOTE:
If the ambient air in the local environment contains a significant amount
of dust and lint, the cleaning frequency must be increased to compensate
for these conditions.
6.20.1
A/C SCAVENGER CLEANING
6.20.1.1
Remove the scavenger hoses and drain any accumulated
moisture. Inspect the hoses for deterioration, then reinstall or
replace it if needed.
6.20.1.2
Remove the safety relief valve housing by unscrewing it in a
counter-clockwise direction.
6.20.1.3
Inspect the O-ring and replace it if needed.
6.20.1.4
Remove the safety relief valve from its housing by twisting it
out in a counter-clockwise direction. The tips of needle nose
pliers can be used to turn the valve. Be careful not to damage
the valve’s fragile disk.
6.20.1.5
Remove any accumulated lint or dust from the valve with a
soft brush. The valve may be further cleaned with a low flow
of Air or Oxygen. The scavenger body can be cleaned with a
moist cloth.
6.20.1.6
Reinstall the valve into the housing, making sure that it is
threaded all the way into the housing and that the plastic
washer is properly seated on its upper surface.
6.20.1.7
Reinstall the valve housing onto the scavenger body, making
sure that the O-ring is properly seated.
6.20.2
AC SCAVENGER TESTING
6.20.2.1
Set the AUTO/BAG valve to the BAG position.
6.20.2.2
Open the APL valve.
6.20.2.3
Occlude the bag mount connector with a Breathing System
Leak Test Device (P/N S010159).
6.20.2.4
Interconnect the inspiratory and expiratory valves or
expiratory port on the ultrasonic flow sensor, if applicable
with a 22-mm hose.
6.20.2.5
Set the Oxygen flow on the anesthesia machine to 8 l/min.
6.20.2.6
Install a scavenger adapter (P/N 4108114) between the 19-mm
hose terminal on the scavenger and the scavenger hose.
(
9
)
Summary of Contents for Narkomed GS
Page 2: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 10: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 40: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 46: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 50: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 52: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 90: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...
Page 91: ...RETURN TO CD ROM TABLE OF CONTENTS RETURN TO THIS MANUAL S TABLE OF CONTENTS ...