4-4
Narkomed GS Operator’s Instruction Manual
Draeger Medical, Inc.
4
Preuse Checkout
c.
Interconnect the inspiratory valve and the expiratory hose terminal
on the ultrasonic flow sensor with a 22 mm hose.
d. Completely open the APL valve by turning the control knob fully
counterclockwise to its stop position.
e. Turn the
SYSTEM POWER
switch to
ON
.
f.
Open the oxygen flow control valve to a flow of 8 l/min.
g. Occlude the bag mount opening and watch for a pressure increase
on the breathing system pressure gauge. This pressure increase
must not exceed 3 cm H
2
O.
Breathing
System Leak
Test
7. Perform a breathing and fresh gas delivery system pressure test. This
test detects leaks from the patient breathing system and fresh gas
delivery system.
To perform the test:
a. Close all flow control valves on the anesthesia machine.
b. Turn the
SYSTEM POWER
switch to the
STANDBY
position.
c.
Turn the vaporizers to 0% concentration.
d. Interconnect the inspiratory valve and the expiratory hose terminal
on the ultrasonic flow sensor with a 22 mm breathing hose.
e. Set the manual/automatic selector valve to
BAG
.
f.
Close the APL valve by turning the knob fully clockwise to its stop
position.
g. Check that the breathing pressure gauge is on 0.
h. Attach the supplied test terminal to the breathing bag mount.
i.
Connect a sphygmomanometer squeeze bulb (available from
Draeger Medical) to the hose barb on the test terminal.
j.
Pump the squeeze bulb by hand until the breathing system
pressure gauge indicates pressure of at least 50 cm H
2
O (not to
exceed 80 cm H
2
O).
k. Observe the pressure drop at the breathing system pressure gauge.
When the pressure is at 50 cm H
2
O, begin counting seconds. The
pressure must not drop more than 20 cm H
2
O in 30 seconds.
Scavenger
System
8. * Verify the performance of the scavenger system.
To test the open reservoir scavenger system:
a. Make sure a 19 mm scavenger hose is connected between the
ventilator relief valve and the scavenger interface.
Summary of Contents for Narkomed GS
Page 2: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 10: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 12: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 20: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 62: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 70: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 76: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 96: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 100: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 104: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 186: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 196: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 202: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 203: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...