Draeger Medical, Inc.
Narkomed GS Operator’s Instruction Manual
9-3
Main Switch Panel
9
System Power Switch
The
SYSTEM POWER
switch on the Narkomed GS has two positions:
ON
and
STANDBY
. In the
ON
position the gas (pneumatic) and electric power
circuits are activated and the green LED indicator adjacent to the switch
is illuminated. In the
STANDBY
position the gas supplies, the monitoring
system, and all electrical power to the machine except the convenience
receptacles and battery charging circuit are deactivated.
NOTE:
The battery charging circuit and convenience receptacles are
active whenever the power cable is attached to an active wall
receptacle, regardless of the switch setting.
Testing the Battery
The backup battery system shall be tested daily. To test the battery:
1. Turn the
SYSTEM POWER
switch
ON
, and wait until the power-on
diagnostic screen is no longer displayed.
2. Remove the machine's power plug from the electrical outlet.
3. Press and hold the
BATTERY TEST
button.
4. If the battery is sufficiently charged, the green
BATTERY TEST
light
illuminates after a short delay.
Usually, a fully charged battery can power the electrical components of the
anesthesia machine for at least 30 minutes in the event of a power failure.
NOTE:
During an AC power loss, the
BATTERY LOW
indicator is
illuminated when the battery reserve approaches depletion.
However, do not rely solely on this indicator for an assessment of
battery capacity. If the backup battery becomes completely
depleted and the machine does not have AC power, the
BATTERY
LOW
indicator light has no source of power and does not function.
Therefore, always remember to perform the pre-use battery test.
To prevent premature battery failure, use backup battery power
only during interruption of primary AC power. Do not start an
anesthetic procedure on the anesthesia machine if the
AC POWER
FAIL
indicator light or the
BATTERY LOW
indicator light is
illuminated.
Summary of Contents for Narkomed GS
Page 2: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 10: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 12: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 20: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 62: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 70: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 76: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 96: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 100: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 104: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 186: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 196: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 202: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 203: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...