Daily Checkout Procedure
b. Check the delivered oxygen concentration. Repeatedly flush the
patient breathing system by pressing the oxygen flush button.
Open the oxygen flow control valve to a flow of 8 l/min and close the
other flow control valves. The oxygen measurement display area
should indicate 97% to 100% oxygen concentration.
Fresh Gas
Verification
18. * Make sure all vaporizers are closed. Open the oxygen flow control
valve to an 8 l/min flow and close all other flow control valves. Sniff the
gas coming from the fresh gas common outlet. There should be no
noticeable odor.
Bain Circuit
Adapter
Verification
19. * Verify that the inner tube of the Bain circuit is intact and not
occluded. First deliver a flow of oxygen to the Bain circuit through the
fresh gas hose. Then occlude the inner tube of the Bain circuit. The
oxygen flowmeter float should drop in response to the occlusion.
As an alternate test, press the oxygen flush button with the Bain
circuit's patient port open to the atmosphere. The high flow of gas
through the Bain circuit's inner tube will draw in gas from the outer
tube. As a result, the breathing bag should deflate. If the breathing bag
does not deflate or it inflates, the fresh gas hose or inner tube may be
improperly connected.
Absorber
System
Verification
20. * To check the absorber system:
a. Check the hose connections in the breathing system.
b. Make sure the fresh gas hose of the breathing system is securely
connected to the fresh gas outlet.
c.
Make sure a 22 mm patient breathing circuit is connected between
the inspiratory valve and the expiratory hose terminal on the
ultrasonic flow sensor.
d. Make sure a 22 mm breathing hose is connected between the
ventilator hose terminal and the manual/automatic selector valve
breathing hose terminal.
e. Make sure a breathing bag of proper capacity and appropriate
construction is connected to the breathing bag terminal of the
breathing system.
f.
Make sure the breathing pressure pilot line is properly connected
between the
BREATHING PRESSURE
interface and either the
absorber quick-connect fitting or the appropriate fitting at or near
the patient Y-piece.
g. Make sure the oxygen sensor and flow sensor are properly
installed.
Summary of Contents for Narkomed GS
Page 2: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 10: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 12: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 20: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 62: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 70: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 76: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 96: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 100: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 104: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 186: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 196: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 202: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 203: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...