Specifications
17
Absorber System
Inspiratory
Valve
Mounting ring nut size . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . M35 x 1
Hose terminal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22 mm male
Expiratory
Valve
Mounting ring nut size . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . M33 x 1
Hose terminal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22 mm male
PEEP Valve
(optional)
Range . . . . . . . . . . . . . . . . . approx. 2–15 cm H
2
O (continuously adjustable)
Breathing
System
Pressure
Gauge
Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . -20 to +80 cm H
2
O
Smallest scale division . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 cm H
2
O
Nominal accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . -20 to +5 cm H
2
O: 3% FS
+5 to +55 cm H
2
O: 2% FS
+55 to +80 cm H
2
O: 3% FS
Mounting ring nut size . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 1/8 x 18
APL Valve
Nominal low flow resistance. . . . . . . . . . . . . . . . . . . . . 2 cm H
2
O at 8 l/min
Hose terminal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19 mm male
Breathing
Bag Terminal
Bag terminal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22 mm male
Oxygen Monitoring
Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10–100 vol % O
2
Resolution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 vol % O
2
Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ±3 vol % O
2
(When calibrated within 18 hours, and
constant temperature and pressure)
Response time
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .≤
25 sec (T90)
Zero drift
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ≤
0.1 vol % O
2
/month
Span drift
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .≤
1 vol % O
2
/8 hours
Temperature error
. . . . . . . . . . . . . . . . . . . . . . . ≤
± 3% of reading (15
°
to 40
°
C)
Sensor service life
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ≥
8 months at 25
°
C,
50% relative humidity, 50% O
2
gas mixture (or
≥
5000% hour CO
2
)
Summary of Contents for Narkomed GS
Page 2: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 10: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 12: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 20: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 62: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 70: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 76: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 96: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 100: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 104: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 186: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 196: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 202: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 203: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...