17-4
Narkomed GS Operator’s Instruction Manual
Draeger Medical, Inc.
17
Specifications
Vaporizers (Vapor 19.1 and 19.3)
Temperature Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . +15–35
°
C
(at normal atmospheric pressure of 760 mmHg)
Flow Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0.25–15 l/min
Maximum Pressure Load . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 150 mmHg
(above atmospheric)
Maximum Angle of Inclination . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
°
Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . approximately 7.5 kg
The following values refer to individual concentration settings when
operated with a continuous flow of air in the range 0.25–15 l/m,
temperature at 22
°
C, and normal atmospheric pressure (760 mmHg).
Halothane
Adjustment range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0.2–5 vol%
Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . ± 0.15% concentration (volume)
or
± 15% (whichever is higher)
Enflurane
Adjustment range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0.3–7 vol%
Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . ± 0.2% concentration (volume)
or
± 20% (whichever is higher)
or
+ 20%/- 30% with flow settings
6.0–15 l/min and handwheel settings
higher than 5.0% volume concentration
Isoflurane
Adjustment range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0.2–5 vol%
Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . ± 0.15% concentration (volume)
or
± 15% (whichever is higher)
Sevoflurane
Adjustment range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0.3–8 vol%
Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . ± 0.2% concentration (volume)
or
± 20% (whichever is higher)
Ventilator
Frequency . . . . . . . . . . . . . . . . . . . . . 1–99, ±1 BPM (in 1 BPM increments)
I:E ratio
. . . . . . . . . Standard range
: 1:1–1:4.5, ±0.1 (in increments of 0.5);
Extended range
: 4:1, 3:1, 2:1
Inspiratory flow. . . . . . . . . . . . . . . . . . . . . . . . . 10–100 l/min (uncalibrated)
Tidal volume . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20–1500 ml, ±100 ml
Pressure limit control adjustment range. . . . . . . . . . . . . . . . 15–120 cm H
2
0
Summary of Contents for Narkomed GS
Page 2: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 10: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 12: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 20: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 62: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 70: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 76: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 96: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 100: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 104: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 186: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 196: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 202: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 203: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...