Draeger Medical, Inc.
Narkomed GS Operator’s Instruction Manual
3-9
Daily Checkout Procedure
3
g. Set the absorber's manual/automatic selector valve to
BAG
.
h. Turn the APL valve control knob fully counterclockwise.
i.
Verify that the suction waste gas disposal system is active.
j.
Adjust the scavenger needle valve to a flowmeter indication
between the two white lines.
k. Close all flow control valves on the anesthesia system.
l.
Occlude the absorber breathing bag terminal and observe the
breathing pressure gauge on the absorber. The gauge should
indicate only a negligible negative pressure (no lower than -0.5 cm
H
2
O).
To test for positive pressure relief:
m. Perform steps a. through d.
n. If the system is equipped with a PEEP valve, turn the PEEP valve
control knob fully counterclockwise to its lowest setting.
o. Adjust the scavenger needle valve to a completely closed position by
turning it fully clockwise.
p. Open the oxygen flow control valve to a flow of 10 l/min and occlude
the absorber breathing bag terminal.
q. The flow of oxygen must now exit through the relief ports located on
top of the canister. The absorber system's breathing pressure gauge
must indicate a pressure less than 5 cm H
2
O.
r.
After the test, adjust the scavenger needle valve to a flowmeter
indication halfway between the two white lines.
To test the scavenger interface for passive systems:
a. Make sure a 19 mm scavenger hose is connected between the
ventilator relief valve and the scavenger interface.
b. Make sure a 19 mm scavenger hose is connected between the APL
valve on the absorber and the scavenger interface.
c.
Check for moisture accumulation in the breathing and scavenger
hoses. Remove any moisture found.
d. Short-circuit the inspiratory valve and the expiratory hose terminal
on the ultrasonic flow sensor with a 22 mm breathing hose.
e. Set the absorber's manual/automatic selector valve to
AUTO
.
f.
If the system is equipped with a PEEP valve, turn the PEEP valve
control knob fully counterclockwise to its lowest setting.
g. Open the oxygen flow control valve to a flow of 10 l/min and occlude
the 19 mm scavenger terminal labeled
EXHAUST
.
Summary of Contents for Narkomed GS
Page 2: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 10: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 12: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 20: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 62: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 70: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 76: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 96: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 100: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 104: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 186: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 196: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 202: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Page 203: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...