Trigenflex R96
Ceiling, Wall, Standmodel
Dr. Mach
Lamps and Engineering
59230001
Edition 01
18.02.2002 / Bak Page 5/23
♦
Voltage measurement at the control board
Please check, whether the power supply is supplying AC or DC.
Voltage measurement at DC
Pay attention to the polarisation during the installation.
If the light intensity control does not function as desired, the PLUS- and MINUS-pole at the power
supply have to be changed.
Voltage measurement at AC
For installation at AC proceed as described below:
•
Preparation
The lamp is already switched on. Set the lamp to
the
lowest light intensity
.
•
Measurement
Measure the voltage at measuring points
3 + 4
. The
voltage must be less than 40V DC.
If this value is exceeded, reset it as described be-
low:
Measuring points 3+4
♦
Voltage setting at the control board
Switch the lamp off.
The position of the control board is shown in the
figure.
Adjust the standard setting by changing the jump-
ers (see arrow).
Three settings enable you to ideally set your lamp
system (J4 and J5).
Take off one jumper and put it into position as
shown figure II. or III.
Switch on the lamp. Set the lamp to the
lowest
light intensity
.
Repeat the voltage measurement at measuring
points
3 + 4
.
If the voltage is less than 40V DC, keep the chosen
setting.
J4 J5
Control board
J5
J5
J5
J4
J4
J4
I.
II.
III.
J5
J4