Trigenflex R96
Ceiling, Wall, Standmodel
Dr. Mach
Lamps and Engineering
59230001
Edition 01
18.02.2002 / Bak Page 4/23
2. Operating the lamp Trigenflex R96
The dielectric filter disk between reflector and splinter protection disk prevents a damaging heating of the
illuminated area.
The lamp may not be used without the dielectric filter.
2.1 Checking the lamp Trigenflex R96 with light intensity control
•
Voltage measurement and setting
Before using the OT-lamps, a voltage measurement and, if necessary, a voltage setting has to be done.
Only in this way a perfect functioning can be ensured.
If the location of the lamp is changed, the voltage measurement and setting must be repeated.
♦
Voltage measurement at the halogen bulbs
To measure the voltage proceed as follows:
Mount the lamp to the ceiling or wall attachment.
Switch on the lamp.
In order to make the measurement remove the
cover
. Remove the screw in the middle of the cover
and take off the cover.
The measuring points for checking the voltage are
in the same position on all Trigenflex-lamps.
The measuring points are shown in the figure.
At measuring points 1 + 2 measure the voltage
applied at the halogen bulbs.
The lamp is set to
the maximum light intensity.
The AC or DC voltage depends on the external
power supply.
Rated voltage 22,8-23,8V.
Screw
Measuring points 3+4
♦
Voltage setting at the halogen bulbs
If the voltage measured is too high or too low, the
value must be adjusted at the power supply.
The setting at the transformer is made by recon-
necting the wires on the secondary side.
Measuring points 1+2
Measuring points 1+2