C L 1 U S E R S G U I D E
D O D T E C H N O L O G I E S , I N C
Rev 10.07
Page 22
5.6 Service Menu
The Service menu is intended for client use but should remain password protected to prevent
unauthorized access. The default Service Menu password is included with this manual. The
password can be changed as needed (see section 5.6.c)
5.6.a Main Menu
Selecting this menu item will return to the main menu (Section 5.4)
5.6.b Optic Calib (Figure 5.26)
This screen is used to manually calibrate the optics which normally only needs to be done
when the system is first configured. Touch the <Start> soft key to start the calibration
process. When the process starts the DAC & Ref values will be set to 0. When the process
completes both values will update with non-zero values indicating that the process is
complete.
Figure 5.26
5.6.c Passwords(Figure 5.27)
This screen is used to change the passwords for both the Setup Menu and Service menu. A
password of 0 will disable the password feature for the specified menu. The default
passwords are included with this manual.
Figure 5.27
Use the arrow keys below the screen to select among the password fields and enter the values
using the procedure outlined in section 5.1.3. The passwords are automatically saved and set
to the value displayed.
5.6.d Optic Config
Selecting this menu item will display a screen of values that may aid service personnel in
verifying the optic system. Under normal conditions the screen is not used. Hit <ESC> to
return to the Service Menu.
WARNING
– Be sure to keep the Service menu password in a safe place. If the
service password is lost you will no longer be able to access the service menu and
may need to restore default values and completely reconfigure the CL1.