C L 1 U S E R S G U I D E
D O D T E C H N O L O G I E S , I N C
Rev 10.07
Page 17
5.5 Setup Menu (Figure 5.14)
The Setup Menu contains the screens necessary to configure the CL1 for operation. The setup
menu is not password protected by default but may be password protected (see service menu).
Figure 5.14
Each selection on the setup menu is detailed below
5.5.a Main Menu
Selecting this menu item will return to the main menu (Section 5.4)
5.5.b Tape Saver (Figure 5.15)
(Default configuration : Disabled)
DANGER:
Make sure the Tape Saver Modes are completely understood before
enabling. Tape Saver modes allow the operator to bypass continuous gas monitoring.
Tape Saver Mode is used to reduce tape advancement during a gas release. Three tape saver
modes are available on the CL1 : Fixed Cycle mode
(Default)
, Single Reading, and
Continuous. Use the upper left soft key to select among the three modes.
Figure 5.15
In Fixed Cycle mode the Minutes entered on the screen is the MINIMUM time each cycle will
take regardless of whether gas is detected or not. The maximum time of the cycle is the
normal cycle time of the CL1 tape being used(normally approx 54 minutes). Enter the # of
minutes (1-54) which will be the minimum time before tape advance. If gas is detected and
the tape becomes saturated in less than the time specified, the system will wait until the cycle
time is complete before advancing. When this occurs the timer on the Analysis screen will
begin counting down the # of seconds remaining until tape advance.
In Fixed Cycle Mode
the conc. Level is not used and will always show a non-zero value.
In Single reading mode the time specified (minutes) is the exact analysis cycle time. At the
start of each cycle the system will take 1 reading and report the concentration(including zero).
The timer will then continue to countdown to zero before advancing to the next cycle.
In
Single Reading Mode the conc. Level is not used and will always show 0.
Current Mode
Current state of Tape
Saver Mode
# of seconds to wait before
tape advance
“Trigger” Concentration
Use this soft key to select
among 3 Modes
Use this soft key to toggle
TSM on and off.