17
SETTINGS MENU
The “Settings Menu” contains several sub-menus that can be accessed
to change functional aspects of the interface. These sub-menus
include:
1. FORCE CALIBRATION
2. pH CALIBRATION
Enter the FORCE CALIBRATION sub-menu to perform the calibration
procedure.
3. FORCE REC. OUTPUT
4. MEASUREMENT RANGE
5. INTERFACE SETTINGS
These are the options that are present when the system is in
Calibrated Mode. Different options will appear when the system is in
Direct Mode.
1. Force calibration
NOTE: everytime a force calibration is performed the
MEASUREMENT RANGE is set to default 200 mN. Set
measurement range after the force calibration.
Touch “SELECT” to initiate the calibration for the chosen chamber
and the text will turn blue. Touching “ENTER” will start the 6-step
procedure for calibrating the chosen force transducer.